GwCT-100 current Transformer ratio Tester Transformer Comprehensive Tester for indutance winding turns Resistance
GWCT-100 automatic transformer testing machine uses PLC control technology.
It can be customized according to the specific needs of customers. This versatile device is capable of simultaneous control of the Impulse Winding Tester, LCR Meter, Precision Miniature Current Transformer Tester, DC Resistance Meter, and Fiber Laser Marking Machine and Uv Online Laser Marking Machine, enabling accurate and efficient testing of multiple processes.
Our state-of-the-art technology ensures that every test is performed in accordance with industry standards, providing users with unmatched accuracy and reliability. This equipment can automatically make judgments, effectively saving time and labor costs, and avoiding possible errors caused by manual judgment.

Impulse Winding Tester
I. Specification:Main technical parameters:
|
Model |
GW-IT15-5KV |
GW-IT15-3KV |
GW-IT15-10kV |
|
|
Output pulse voltage |
100V~5000V 10V step 5%±10V |
100V-300V10V step 5%±10V |
300V ~ 10000y step 5%±10V |
|
|
Inductance test range |
>10μH |
|||
|
Maximum pulse energy |
0.25J |
|||
|
Display Waveform display area Display content |
800×480 dot matrix graphic LCD 650×250 dot matrix Enlarged display, setup parameters, standard and test waveforms, comparison results, file information, etc. |
|||
|
Waveform sampling |
Sampling rate: 100MSa/s~100kSa/s, total 10 steps Resolution: 8bits Sampling point: 6500bytes |
|||
|
Standard waveform sampling |
Continuous cycle, single cycle, single sample (up to 32 times sample average function) |
|||
|
Input impedance |
10MQ |
|||
|
Measurement speed |
Maximum 15 times/sec |
|||
|
Number of pulses |
Test pulse |
1~32 times programmable |
||
|
Degaussing pulse |
0~16 times programmable |
|||
|
Display waveform measurement |
Voltage, Time, Frequency |
|||
|
Trigger mode |
Internal Internal/Manual Man μ AI (with foot control Foot)/External/Bus |
|||
|
Comparison Methods |
Area Size Comparison ,DiffZone Comparison Corona Comparison,PhaseDiff Comparison |
|||
|
Area Repeatability |
±1% |
|||
|
Area Difference Repeatability |
±1% |
|||
|
Judgment output |
PASS, FAIL display, LED display, buzzer alarm |
|||
|
Alarm volume selection |
Long Treble, Long Bass, Single Short Tone, Double Short Tone, Off |
|||
|
Memory |
Internal: 200 groups of files External USB |
|||
|
Interface |
HANDLER(START,STOP,PASS,NG,BUSY,EOC.),RS232C IEEE488(Option),USB DEVICE,USBHOST |
|||
|
Volume and weight |
400mmx132mmx300mm, weight 8.0Kg. |
|||
II.Brief description
GW-IT15 series pulse coil tester is specially designed for the testing needs of wire-wound components, the use of a high-voltage charging of a small capacitor (low test energy) and the coil to be tested to form a parallel resonance of the RLC, by the attenuation of the oscillating waveforms, through as high as 100Msps (can be customized to 200Msps) and the precision of the sampling processing and analytical techniques, can be tested to the coil from the body of the insulation is bad (turn-to-turn short-circuiting, short circuit between layers, insulation performance degradation, etc.), so that component manufacturers and users can more effectively for product quality control. Short circuit between turns, short circuit between layers, insulation performance degradation, etc.), so that component manufacturers and users can be more effective for product quality control.
GW-IT15 series instrument measurement speed, test waveform repeatability, can be widely used in motors, frequency transformers, relays, switching power supply transformers, demagnetization coils, etc. by the enameled wire wound components; GW-IT15 series can be applied to automated test system to greatly improve the test efficiency.
III.Ordering Information
GW-IT15-3KV Pulse voltage output: 100-3000V
GW-IT15-5KV Pulse voltage output: 100-5000V
GW-IT15-10KV Pulse voltage output: 100-10000V
GW-IT15S-22 Channel Test, Pulse Voltage Output: 100-5000V
GW-IT15S-44 Channel Test, Pulse Voltage Output: 100-5000V
GW-IT15S-88 Channel test, pulse voltage output: 100-5000V
GW-IT15S-XX10 channels-100 channels, pulse voltage output:100-10000V, can be customized according to customer's requirements.
IV.Advanced Pulse Test Principle
The so-called"pulse testing of wire-wound components " is a non-destructive, high-speed, low-energy voltage pulses applied to the test
piece to analyze and compare the waveforms of the good and defective test waveforms for the purpose of judging the GW-IT15 series of large screen, ultra-high sampling rate, high-performance processing technology, so that the customer intuitively distinguish the difference between defective and good products. The GW-IT15 series has a large screen, high sampling rate and high performance processing technology.

V.Powerful corona test function
Can be customized up to 200Msps high-speed sampling rate, accurate extraction of corona information, and is equipped with corona display function, corona display at a glance, the performance of the product to do to prevent problems before they occur, greatly improving product quality.

VI.Unique destructive testing capabilities
Destructive test is a test mode that starts from the starting voltage, takes the step voltage as the voltage step, ends with the termination voltage and combines with the comparator for comparison, which is used to judge the limit voltage withstand value of the tested parts, which can help customers to know the voltage withstand value of their own products and serve as the reference for setting the test voltage.
VII.Expanded dual-channel, quad-channel testing capabilities
GW-IT15S-2 dual-channel, GW-IT15S-4 four-channel can meet the customer on the Y-type winding (three-phase four-wire), three-phase three-wire winding test needs, motor winding test once, to avoid multiple switching test troubles, testing fast, efficient and cost-effective, widely praised by customers.
It can also be equipped with an external multi-channel scanner box, such as a typical 8-channel scanner box.
VIII.FEATURES:
65k color 7-inch TFT 800*480 HD display, Chinese and English optional operation interface
100Msps waveform sampling rate, 6500bytes memory depth (200Msps can be customized).
High-speed testing, up to 15 times/sec.
Four waveform comparison judgment methods: area, area difference, corona and phase difference
Excellent voltage stability, test repeatability, to ensure the stability of the test
Powerful corona analysis and extraction function, digging out the potential insulation problems of the product.
Automatic saving of instrument parameters and power-on file loading function.
Waveform vertical zoom, horizontal zoom and move function, easy to observe details
Sample averaging function, up to 32 times of standard waveform averaging processing
Automatic standard acquisition mode, automatically select the appropriate sampling rate
Destructive testing for you to select the appropriate test voltage
Demagnetization pulse application to ensure consistent test waveforms Firmware version update via USB HOST.
Foot-control interface for quick and easy measurements.
Extensive interfaces for Handler, RS232C, USB Device, USB Host, and GPIB (optional).Provides up to 4 channels of testing, can test any three-phase motor delta and star connection, single-phase motor main and vice winding, main and vice winding, main and vice + main and vice winding, to meet the full range of motor turn-to-turn testing, can also provide 4 windings, and can also provide 4 winding testing.It can also provide scanning test function of 4 windings and winding comparison test function.
LCR Meter
Specification:Main technical parameters:
|
model number |
GW-IT16B |
GW-IT17C |
||
|
Test Frequency |
50Hz-200kHz 37 frequency points |
50Hz,60Hz,100Hz,120Hz,1kHz, 10kHz,20kHz,40kHz, 50kHz, 100kHz,total 10 points |
||
|
test parameter |
C.L,X,B,R,G,D,Q,ESR, Rp. θ .|Z|,|YI ,P |
C,L.X,B.R,G,D,Q,ESR,DCR,TURN.Rp,θ ,|ZI,IYI,P |
||
|
test level |
10mV-2Vrms |
|||
|
Fundamental Accuracy |
0.05% accuracy table prevails |
|||
|
test speed |
Fast: 50 times/sec, medium 20 times/sec, slow 6 times/sec (frequencies above 1kHz) |
|||
|
Display range |
|ZI,R,X |
0.01MΩ~99.9999MΩ |
||
|
YI.G,B |
0.00001μS~99.9999S |
|||
|
C |
0.00001pF~9.99999F |
|||
|
D |
0.00001~9.99999 |
|||
|
DCR |
0.01MΩ~99.9999MΩ |
|||
|
L |
0.00001μH~99999.9H |
|||
|
Q |
0.01~99999.9 |
|||
|
θ(RAD) |
3.14159~3.14159 |
|||
|
θ(DEG) |
-179.999°~ 179.999 ° |
|||
|
△% |
-99.999%~99.999% |
|||
|
Turns-Ratio |
1:0.01-100:1 |
|||
|
Signal Source Output Impedance |
30Ω ,100Ω |
|||
|
Internal DC bias voltage |
0V,2V |
0V,1.5V,2V |
||
|
Ranging method |
Automatic, Hold |
|||
|
Trigger mode |
Internal, Manual, Auto, External, Bus |
|||
|
Calibration Function |
Open/Short, Full Frequency Zero, Spot Frequency Zero |
|||
|
Equivalent circuit |
Series, Parallel |
|||
|
Comparator |
10-position selector,,BINO~BIN9,NG,AUX |
|||
|
PASS,HI ,LOW(PASS/FAIL LED display) |
||||
|
Interfaces |
RS232C,Usb Host,USB Device,Handler,GPIB(Option) |
|||
|
Temperature and Humidity |
0ºC~40ºC ,Relative Humidity≤75%. |
|||
|
Power Requirements |
Voltage: 100~120 Vac or 198~242 Vac |
|||
|
Frequency: 46~64 Hz, power greater than 50 VA |
||||
|
Volume(WxHxD) |
310mmx115mmx315mm |
|||
|
Weight(Net) |
4.5Kg |
|||

FEATURES:
1. Highly comparable with international brand LCR bridge test data (Keysight,WK)
2. Keyboard lock function, power-down data protection function
3. List scanning function, open circuit, short circuit, load correction function
4. Simplified Chinese and English operating languages
5. Test data, test conditions saving (USB flash disk or internal)
6. Test voltage, current automatic balance adjustment function (ALC)
7.50 sets of test files inside the instrument and 500 sets of test files on USB flash disk can be recalled at any time
8. U disk can support the format FAT16, FAT32 file system
Brief statement
GW-IT16B/GW-IT17C digital LCR bridge is a new generation of impedance tester with color TFT display screen, its basic accuracy of 0.05%,
50Hz~200kHz frequency range can be used to meet the incoming inspection, shipment inspection, to provide product quality assurance.
Ordering Information
GW-IT16B:50Hz-200KHz LCR Digital Bridge
GW-IT17C:50Hz-100KHz LCR digital bridge with lap count function.
Precision miniature current transformer tester (current three-piece set)
I.Summarize:
This set of instruments is used for error testing of miniature current transformers. It adopts the principle of comparison method. The whole set of test instrument includes electric ascender, miniature standard current transformer, miniature transformer calibrator and the supporting power cord, connecting cable and so on. Integral good, easy to operate. The instrument is designed with reference to the miniature current transformer JB/T10665-2006 standard, which meets the needs of error measurement of miniature current transformers of 0.02 level and the following levels.
The new calibrator is superior to the characteristics of similar products:
1. Test response speed is greatly improved. At the same time increase the fast and slow speed switching function, to meet the different requirements of the test speed of the customer.
2. Touch screen display and control, simple operation and intuitive display.
3. Increase the alarm setting function switch, open and log in to set their own data collection percentage points, angular difference and than the difference of the alarm prompt range. According to the set accuracy range, the error will be judged whether it is qualified or not, and the red color will be displayed when it exceeds the tolerance and accompanied by an alarm tone.
4. At the same time to increase the loudness alarm and switch,according to the user site environment to open the loudness alarm, while adjustable size. Meet the site environment of noisy users.
5. Increase the automatic collection function, set the percentage point to be collected by yourself, enter the data collection interface, save the data automatically after meeting the test requirements, and the data can be exported to excel document with USB flash disk.
6. Add control output function, according to the setting to judge whether it is qualified or not and then give the corresponding signal, it can be connected with the automatic platform to realize automatic control.
7. 485 communication can be connected to the computer or the console to realize automatic control.

II.Main technical indicators
1. Primary rated current:
a: 100, 75, 30, 25, 20, 15, 10, 6, 5, 3, 2.5, 1.5, 1, 0.6, 0.3
2. Secondary rated current:
mA: 60, 50, 40, 30, 25, 20, 50/3 , 15, 10, 7.5 , 20/3 , 6, 5, 4, 10/3, 3,2.5 , 2, 1.5 , 1
3. Multi-range standard current transformer accuracy: 0.005 level.
4. Measuring line accuracy: display value error: ≤ ± 2% phase error: ≤ ± 2% radian
5. power supply: maximum output power: 300VA, with fine-tuning.
6. load resistance: 0Ω , 5Ω , 10Ω , 15Ω , 20Ω , 25Ω , 30Ω , 40Ω , 50Ω , 60Ω ,70Ω , 75Ω , 80Ω , 100Ω , 150Ω , 200Ω , 250Ω , 500Ω , 1KΩ , 1.5KΩ , 2KΩ , external.
7. Conditions of use:
(1) Power supply: utility 220V, 50Hz and 60Hz
(2) Use of the environment: temperature 10------40 ºC; humidity ≤ 80%
III.Structural description
This instrument is a vertical three-box structure.
The lower box is the booster, i.e. power supply box. The output current has 1A, 5A and 100A gears. According to the primary current of the
measured transformer to choose the output gear of the booster. It is also equipped with coarse and fine regulator for current adjustment.
It is strictly prohibited to use the power supply box to exceed the current limit, otherwise it may cause permanent damage to theinstrument!
Micro standard current transformer. The panel has a primary current terminal, secondary current transfer switch; the upper box for the
measurement error line box: miniature transformer calibrator. On the panel there is a touch screen, set the function and display data; the measured transformer secondary terminals K1,K2; the measured transformer load (resistance) changeover switch, external load terminals. When the secondary load of the measured transformer is a load other than the designed load, the external load terminal can be connected to the required load value. In this case, the load knob should be placed in the external position.
The connections on the front and back of the three boxes are shown in the following diagram.
Picture a Picture b
IV.Usage
1. according to the above figure after connecting the cable line,
according to the measured transformer range, select the corresponding standard transformer range.
2. the lower box regulator is placed in zero position, turn on the power. Adjust the regulator to the required current value, read f, δ error.
3. the specific method of measuring the number of turns, the primary current is indicated in amperes, the secondary current is indicated in
milliamps. First of all, according to the size of the primary current of the measured transformer in the standard selection of the corresponding wiring jack, and then the formula: " primary rated current * primary turns * 1000 / secondary turns" to get the size of the output secondary current, the secondary switch to the corresponding gear calculated.
For example: a current transformer primary current of 3A, primary turns for 1 turn, secondary turns for 50 turns, then the primary jack should be selected 3A gear, calculated secondary switch gear 3 * 1 * 1000 / 50 = 60mA. secondary switch should be dialed to 60mA gear.Other small transformer turns measurement and so on. Based on the above method, we can measure many kinds of turns of the transformer, the fundamental approach to calculate the turns of the corresponding relationship, if the primary 40A1N
current transformer, secondary turns for 1000N, we conclude that the secondary current should be 40mA, but the standard primary wiring jacks do not have 40A, we will be commuted to the other corresponding gear to measure, only need to satisfy the corresponding relationship between the number of turns can also be used. Can be, for example, with the primary 30A (see the maximum range), secondary 30 mAto measure, but also with the primary 20A, secondary 20 mAto measure, but pay attention to the current operating range.
V.Precautions
1. access to the measured transformer, should be connected to the secondary first, and then connected to the primary. When removing the transformer, the primary should be removed first, then the secondary. And pay attention to whether the polarity is correct or not.
2. When measuring the error, the grounding wire of the instrument must be firmly grounded in the lower left part of the rear panel of the instrument, see Picture b.
Note: Please pay attention to the power supply access, L,N (fire wire, zero wire), must not be connected to the opposite.
VI.Touch screen operating instructions
1, the touch screen boot 3 seconds automatically enter the main interface, as Picture 1.
Picture 1
When the communication is normal, it displays the real-time data of the percentage meter, ratio difference and angle difference. At the same time, it displays the wiring status of the transformer, if the transformer is normally wired, it displays "working normally" ; if the transformer is incorrectly wired, it will prompt the secondary open circuit/polarity error/measurement range error.
Click " Menu" and " Description" to enter the corresponding interface;
"Tone on/off" button to switch on/off;
" Unit of Angular Difference" button to switch the unit of angular difference to be displayed in minutes or %;
" Digit +" button increases the display digit by 1 digit; " Digit -" button to decrease the display digit by 1 digit;
"Speed +" measurement data stabilization time is 0. 1S, long press 2S switch; 'Speed -' measurement data stabilization time is 0.5S, long press 2S switch;
2,Menu
As in Picture 2, click " Menu" to enter the corresponding interface.
Picture 2
3,Alarm Setting
As in Picture 3, you can set the accuracy of the transformer, as well as the percentage range of the measured transformer, and the upper and lower limits of the angular and specific difference errors.
Picture 3 Picture 4
When you use it for the first time, first turn on the switch of alarm function, at this time, the upper and lower limits of percentage, angular difference and ratio difference are all 0. Click "Administrator Login " to enter the setting mode and log in as the person in charge (password 1), then press the"Restore Initial Values " button to set the upper and lower limits and the initial values of percentage, angle difference and ratio difference. Initial value. According to the accuracy of the transformer, press the corresponding accuracy button at the bottom of the screen to display the upper and lower limit values of ratio difference and angular difference under the accuracy.When the percentage test range needs to be modified, click on the percentage position to enter the percentage meter setting interface, click on the measurement point position to directly input the desired measurement point, and at the same time, set the upper and lower limit ranges of the point, as in Figure IV. If the upper and lower limits of angular difference and specific difference need to be modified, click the corresponding position to modify, and finally press the "Confirm Modification " button. Return to the main test interface, adjust the regulator, the collected ratio difference and angle difference value exceeds the set error range, then the data display red alarm, at the same time, the buzzer emits an alarm tone,"Tone Switch "can turn off the alarm tone. Meanwhile, the loud alarm tone can be adjusted by the volume knob on the front panel.
Click "Current " button, current CT part of the red box prompts, click "Voltage " button, voltage PT part of the red box prompts.
4,Measuring point setting
As shown in Picture 5, set the alarm prompts and data saved over the difference between the measurement point, selected green prompts, click on the "data acquisition" to enter the automatic acquisition and storage interface, as shown in Picture 6, from 0 slowly adjust the regulator to the maximum measurement point, the left side of the data is automatically collected, if beyond the scope of the setup of the red and the alarm prompts, the selection of all the measurement points collected! After finishing, the percentage meter exceeds the maximum measurement point, prompting the data to be saved successfully, the sample number is automatically added 1, the number needs to be recounted every time the power is turned on, and the number can be set by yourself according to your needs. If there is a missed measurement point, it will prompt the data collection is incomplete and re-measurement.
After the data is saved successfully, it can be queried through the history, and it needs to be exported directly by clicking the data exported to U disk, as in Picture 7.
Picture 5 Picture 6 Picture 7
5,485 Setting
As in Picture 8, you can modify the 485 address, "OK" button to save and return to the '"485 Settings" interface.
Picture 8
6,Calibration
Enter the calibration interface, such as Picture 9, the "secondary range"switch to "calibration" file, K1, K2 terminals are disconnected, the primary line in series with the standard ammeter, monitor the primary current value; adjust the ascender knob, so that the standard ammeter shows 5A, at this time!Percentage meter, ratio difference, angle difference should be displayed 100%, -1.000%, 0.00, if the gap between the displayed values is large, press the "Calibration" key to calibrate the instrument;
Picture 9
Press "Save" key to save the calibration data, if you do not need to save the calibration data directly press " Return" .
VII.485 Serial Communication
(I), serial port output format
Baud rate = 9600, 8 data bits, 1 stop bit, no parity.
(II), COM1 port 485 communication protocol
Execute Modbus protocol, the upper computer is the master, read and write to the touch screen.
1,Sending code format: station number + function code + start address + number of registers + CRC checksum Receiving code format: station number + function code + number of bytes + register value + CRC checksum
Station number: 0X01 Function code: 0X03
Starting address: 100 (2 bytes)
Number of registers: 8 (2 bytes)
Number of bytes: 16 (1 byte)
Register value: Percentage meter (4 bytes) + Ratio difference (4 bytes) + Angle difference (4 bytes) + Operating status (2 bytes) + Communication normal mark (2 bytes), where the data high byte is in the front and the low byte is in the back.
CRC checksum: low 8 bits + high 8 bits
Remarks: working state: indicated by 0-10, 16-bit positive integer 0: working normally;
1: out of the applicable range, do not continue to raise the test current;
2: the transformer range error;
3: Transformer polarity error;
4: Transformer polarity error, range error;
5: secondary open circuit of the measured transformer;
6: instrument failure, can not work;
7: serious wiring error, do not continue to raise the test current; 8: primary current exceeds the equipment limit;
9: error qualified;
10: error beyond the setting range.
2,The upper computer reads the current measured value.
Register address 100[4 bytes]: percentage scale (unit %, 32-bit floating point number)
Register address 102[4 bytes]: ratio difference value (unit %, 32-bit floating point number)
Register address 104 [4 bytes]: angular difference value (unit %, 32-bit floating point number)
Register address 106 [2 bytes]: operating status (16-bit positive integer)
Register address 107 [2 bytes]: communication normal mark, cycle +1, range 1-9999, (16-bit positive integer)
3,Local station number: according to the requirements of the host computer, the instrument interface operation change, range 1-240, the default station number is 1.
(III) COM2 port 485 communication protocol
The instrument sends data through the serial port after each measurement. It supports Modbus protocol or free serial port, and the free serial port intercepts the 8th to 23rd bit of data for display after receiving.
1. Communication code format: station number + function code + start address + number of registers + number of bytes + register value + CRC checksum
Station number: 0X01
Function code: 0X10
Start address: 100 (2 bytes)
Number of registers: 8 (2 bytes) Number of bytes: 16 (1 byte)
Register value: working status (2 bytes) + percentile meter (4 bytes) + ratio difference (4 bytes) + angle difference (4 bytes) + communication normal mark (2 bytes), where the high byte of data is in the front and the low byte is in the back.
CRC checksum: low 8 bits + high 8 bits
Remarks: Measurement status: indicated by 0-10, 16-bit positive integer
0: working normally;
1: out of the applicable range, do not continue to raise the test current;
2: Wrong transformer range;
3: Transformer polarity error;
4: Transformer polarity error, range error;
5: secondary open circuit of the measured transformer;
6: instrument failure, can not work;
7: Serious wiring error, do not continue to raise the test current;
8: Primary current exceeds the equipment limit;
9: error qualified;
10: the error exceeds the setting range.
Percentage meter: unit %, 32-bit floating point number;
Ratio difference table: unit %, 32-bit floating point number;
Angular difference meter: unit %, 32-bit floating point number;
Communication normal mark: cycle +1, data range 1-9999, (16-bit positive integer)
2,Example of communication data
01 10 00 64 00 08 10 00 00 3F 7F BE 77 BD FC B9 24 3F 2D CC 64 00 09 91 C8
Indicates: operating status 0 working normally, percentage meter 0.999%,ratio difference meter -0.1234%, angle difference meter 0.6789%, communication normal mark 9.
3,Contents corresponding to Modbus standard protocol address.
Register address 100 [2 bytes]: operating status (16-bit positive integer)
Register address 101-102 [4 bytes]: Percentage table (unit %, 32-bit floating point number)
Register address 103-104 [4 bytes]: ratio difference value (unit %, 32-bit floating point number)
Register address 105-106 [4 bytes]: angular difference value (unit %, 32-bit floating point number)
Register address 107 [2 bytes]: communication normal flag, cycle +1, range 1-9999 (16-bit positive integer)
(iv) Pin Definitions
VIII.Control output
There are control output terminals on the rear panel, according to the alarm setting range to read whether the error is qualified, the working status prompts "error qualified " that is, qualified output terminals have signal output, the working status prompts "error exceeds the setting range "that is, unqualified corresponding terminals have signal output. Output.
IX.Attachment
1,Power cord: two three-core;
2, standard transformer and calibrator connection line: 1;
3, primary connecting line:
16mm2*0.2m one article;
16mm2*0.8m two (socket);
16mm2*0.8m two strips (gun head with 2 crocodile clips);
4,Secondary connecting line: 1 shielded line (with 2 crocodile clips);
5,Ground wire: 2.5mm2*0.4m two, 2.5mm*0.8m two
6,Communication line: 1 piece;
7, a manual.
DC Resistance Meter
Brief description:
GW-RM13 series DC resistance tester is a color TFT display screen, the basic accuracy of 0.05%, can be used for coil, inductor, transformer, voice coil, conductive film (cloth),heating core, contact resistance, fuse, relay, printed board wire resistance, welding resistance, connector resistance, metal detection, etc., can also be used for automated test equipment
Performance feature:
1.0.05% test accuracy, fast test speed (40 times/second)
2. Whether qualified LED light indicator, different sound prompts, different color prompts
3. Test end contact check function (GW-RM13) and HANDLER output signal
4. Save the test adjustment, 50 inside the instrument, 500 sets of test files on the U disk
5. Keyboard lock function, shutdown automatically save the test conditions, no need to reset next time
6. Flexible interface function, according to the requirements to achieve data recording, saving, processing
7. Minimum resolution: 1μΩ , 6-bit data display, good stability
8. Temperature compensation function can be customized
9. Direct reading value and percentage error display
10. Patented technology, you can upload the results to your computer without installing software
Technical parameter:
|
Model |
GW-R M13 |
GW-RM13HM |
GW-RM13S4 |
GW-RM13A |
GW-RM13B |
|
Test signal |
DC |
||||
|
Basic accuracy |
0.05%+2 words (3M range accuracy of 0.2%) |
0.05%+2 words |
0.1%+2 words |
||
|
Test speed |
Fast: 40 times/s, medium: 15 times/s, slow: 6 times/s |
||||
|
Range mode |
Automatic, save |
||||
|
Trigger mode |
Internal, manual, automatic, external, total circuit |
||||
|
Test current |
1A/100mA/10mA/1mA/100μA/10μA/1μ A |
100mA/10mA/1mA/100 μA/10μA |
1A/100mA/10mA/1mA/100 μA |
||
|
measuring range |
30mΩ/300mΩ/3Ω/300Ω/3KΩ/30KΩ/300 KΩ/3MΩ |
300mΩ/3Ω/300Ω/3KΩ/ 30KΩ/300KΩ |
30mΩ/300mΩ/3Ω/300Ω/3 KΩ/30KΩ |
||
|
Zero clearing function |
short-circuit |
||
|
Comparator |
BIN0-BIN3, NG,Pass,Fail PASS,HI ,LOW(PASS/FAIL LED display),Buzzer cue |
||
|
Port |
GPIB(optional) , RS232C,UsbHost, USB Device,Handler |
||
|
Memorizer |
Internal 50 groups, external U disk 500 groups of files |
||
|
Temperature and humidity |
0°-40° , relative humidity ≤75% |
||
|
Power requirements |
Voltage: 100-120Vac or 198-242Vac Frequency: 46-64Hz, power greater than 50VA |
||
|
Volume (W*H*D) |
310*110*315mm |
216*87*300mm |
216*87*300mm |
|
Weight |
4.5kg |
3.2kg |
3.2kg |
Fiber laser marking machine
Fiber laser marking machine model advantages:
Many innovative technologies of GW-FLM30W series products bring users unparalleled system performance
1. Fully enclosed laser cavity design, dust and no frost. Long life, low failure rate;
2. Innovative full shield laser outer cavity design, no lead exposed, suitable for industrial production environment, anti-interference;
3. Full air cooling refrigeration, high control accuracy, good laser stability, 24 hours continuous operation at room temperature;
4. New fiber laser source, small current power supply, long service life of pump module;
5. Modular structure design, convenient system integration and equipment maintenance;
6. Easy to learn operation software, so that users can easily get started.
Fiber laser marking machine main features:
The product has the characteristics of short pulse, excellent beam quality, high precision, peak power, high modulation frequency and so on. So in metal etc
The processing application of the material surface can significantly reduce the thermal melting effect and achieve perfect marking effect. Not just on metal surfaces Line marking, also suitable for non-metallic materials, such as ABS, nylon, PES, PVC, polycarbonate and so on.
At present, this series of products has been widely used in metal, electronics, computer peripherals, communications, electrical appliances, watches, glasses, jewelry,Clothing, gifts, instruments, trademark signs, automobile manufacturing, ship manufacturing and aerospace industry and many other industries of the product table Surface text, symbols and other marking and micro processing. A rotating work belt is attached for marking circular surfaces, such as ring protection Warm cup, etc., can be four-axis linkage marking.
Technical parameters of fiber laser marking machine:
|
Continuous output power |
≥30W |
|
Laser type |
RC optical maser |
|
Repetition rate |
20-100KHz |
|
Laser type |
Optical fiber 1064mm |
|
voltage |
220V |
|
Cooling system |
Built-in air cooling |
|
Adaptive material |
All metal materials and more than 85% plastic materials |
|
Laser mode M2 |
<1.5 |
|
Marking depth |
0.001-0.5mm(depends on material) |
|
Minimum line width |
0.02mm |
|
Laser spot diameter |
0.015mm |
|
Minimum character |
0.015mm |
|
Repeated positioning accuracy |
0.002mm |
|
Marking range |
150*150mm |
|
Positioning mode |
Red light positioning, focusing |
|
Environmental requirement |
Temperature: 0-40 ° C Humidity: 0-95% |
|
Overall power consumption |
500W |
|
Maximum speed |
20000mm/s |
|
weight |
65kg |
|
Maintenance requirement |
1 year after-service |
Software control system functions:
A new generation of high-speed data processors with newly upgraded control software, from the precision generation of machining graphic signals to marking.High-speed real-time processing of control signals creates a model for high-end applications in the industry.
1. Support dot matrix and vector marking
2. Support file formats generated by commercial and engineering drawing software;
3. Rich graphics editing functions;
4. Powerful circle and text editing functions
5. Support the font and editing of all languages in the world;
6. Directly call Type, True Type, AFM, CID, CFF, SFNT, PRF and other fonts;
7. The marking task can be arbitrarily decomposed, freely arranged, combined and adjusted;
8. Graphics merge processing function;
Import and export files
1. Import files support PLT and DXF files;
2. Support bitmap files;
General and special bar code, two-dimensional code
1. Support standard industry code, 2/5 code, ASCII code, 128 code, EAN13 code; 2.EAN8 code, UPCA code, UPCE code marking;
3. Support DATAMATRIX, Qr code, PDF417 code, 16K code;
Uv online laser marking machine
Uv laser marking machine model features:
The beam quality is high, the spot is very small, and the ultra-fine marking can be realized. Near-perfect marking quality: 355nm output wavelength reduces the thermal impact on the workpiece;
High precision marking head with galvanometer, fine marking effect and repeatable processing;
High precision and fine light spots ensure perfect marking results; Marking process is non-contact, marking effect is permanent;
Uv laser marking machine has a large range of application;
The thermal affected area is very small, there will be no thermal effect, and the material will not be deformed or burnt;
Fast marking speed and high efficiency; The whole machine has stable performance, small size and low power consumption. It is more suitable for processing materials with large thermal radiation reaction;
Can cooperate with the production line, automatic loading and unloading, automatic material in and out;
Suitable for marking most metal and non-metal materials;
Main advantages of UV laser marking machine:
Ultraviolet laser because the focus spot is very small, and the processing heat affected zone is small, so it can be ultrafine marking, special material marking, is the marking effect has higher requirements for customers must choose products.
Uv laser in addition to copper material, almost suitable for processing all materials.Flexible and convenient operating system: humanized operation process, good stability of equipment operation;
Special control software can be compatible with AutoCAD, CorelDRAW, Photoshop and other software output; It can realize the automatic arrangement and modification of text symbols, graphic images, bar codes, two-dimensional codes, serial numbers, etc.
Support PLT, PCX, DXF, BMP, JPG and other file formats, can directly use TTF font; The laser head can flexibly adjust the printing Angle and height position according to the actual position of the product.
Technical parameters of UV laser marking machine:
Laser wavelength: 355nm
Pump mode: semiconductor end pump
Beam quality: M2<1.2
Laser power: 5W
Laser repetition frequency: 10KHZ--100KHZ
Marking range: 180mm×180mm
Marking speed: ≤12000mm/s
Minimum character: 0.1mm
Repeated positioning accuracy: 5um Cooling method: water cooling
Power supply: 220V/50HZ/2KVA
Options and accessories: working platform, laser, vibrating lens, jig
Uv laser marking machine application and market:
Mainly used in the high-end market of ultra-fine processing, medicine, cosmetics, video and other polymer materials packaging bottle surface marking, the effect is very fine, marking and cleaning firmly, better than ink coding and pollution-free;
Flexible pcb board marking, scribing;
Silicon wafer microhole, blind hole processing; LCD liquid crystal glass two-dimensional code marking, glass utensils
Surface drilling, metal surface coating marking, plastic keys, electronic components, gifts, communication equipment, building materials and so on
Grewin Industrial Group Co., LTD, founded in 1993, is one of China's largesttoroidal winding machine manufacturers.
Widely used in toroidal transformer, current transformer, sensor, inductor manufacturing and other related industries.